Aeration Sewage Treatment

Download (581 KB)

Introduction

When used to measure the dissolved oxygen in an aeration tank, sensors tend to quickly become contaminated and need frequent cleaning. A cost-effective solution to this problem has finally been found with the development of the Yokogawa DO402 Dissolved Oxygen Converter system. With this system, the sensor is mounted in a float ball holder that protects it from contamination and enables continuous maintenance-free use for extended periods.

Sewerage Treatment Procedure

Application

Float HoldersSewage that flows into a treatment plant is sent to a primary sedimentation tank to remove the garbage and sludge. Thereafter, it is routed to an aeration tank where activated sludge (sludge containing bacteria, protozoa, and other microorganisms) is added and air is blown into the mixture to agitate it. The microorganisms in the activated sludge thrive on the oxygen and cause the organic substances in the sludge to proliferate and the generated solids to be precipitated.

A shortage of dissolved oxygen (DO) will kill the microorganisms, so the DO concentration in the aeration tank must be properly controlled. A DO analyzer is used to continuously measure the DO concentration.

The spherical float stays relatively clean because contaminants are washed off by the surface-level flow of water. The sensor diaphragm is kept relatively free of contamination as it is positioned at the same height as the top portion of the float.

  • Continuously and stably measures dissolved oxygen concentration
  • Reduces operating costs
  • Eliminates the need for manual cleaning

Product Recommendations

Detector: DO30
Float holders: FD30, PB30
Converter: DO202 DO402

Field Data

DO Readings Data

DO Readings Data

No effects of contamination after four months of continuous operation

We conducted a test with two DO30G sensors that were equipped with the PB350G Angle Float Ball Holder. During the test, no maintenance was performed on one of the DO30G sensors, while the other DO30G sensor was cleaned every two weeks and calibrated monthly. Over the entire duration of the test, the DO30G sensor that did not receive cleaning showed minimal contamination. Furthermore, its measurements did not deviate appreciably from those of the sensor that was regularly maintained.

Precautions:

  • The measuring point must be at a location where air bubbles are not a frequent occurrence.
  • To prevent the sensor diaphragm from being damaged by solids in the sample liquid, the arm of an angled float holder should be pointed in the downstream direction.
  • For more information contact your local Yokogawa Analytical Marketing Department.

Industries

  • Pharmaceutical

    Under continual pressure to increase profitability, maintain government compliance, and meet emerging market opportunities, the pharmaceutical manufacturing industry faces unique challenges that require unique solutions. As one of the world’s leading industrial automation suppliers, Yokogawa is poised and prepared to deliver those solutions, creating individualized lean manufacturing techniques that deliver consistent, measurable results.

    See More
  • Water & Wastewater

    All water, regardless of its source, has value. However, with water resources being finite, a focus on "One Water" is necessary to maximize every drop within the water system, ensuring a stable supply of clean and safe water.  As we consider current and future needs, planning and implementation are critical for long-term resilience and reliability, meeting both community and ecosystem demands. Yokogawa believes water is a human right for all. Thus, contributing to a sustainable water cycle is one of our Sustainable Development Goals (SDGs).

    See More