Yes, by setting the range rating mode to Scale.
For example, with a K thermocouple and measured values with an input span of 0.0°C–100.0°C, to lower all temperatures by 1°C, enter the following settings.
- Mode: Scale
- Range: TC
- Type: K
- Span lower limit: 1.0
- Span upper limit: 101.0
- Scale lower limit: 0.0
- Scale upper limit: 100.0
- Unit: °C
Related Products & Solutions
-
Button Operated DX1000/DX2000
The Daqstation DX1000/2000 series is a data acquisition system with display. Its HMI supports custom graphics for industry applications, audit trail and advanced security conforming to FDA 21CFR Part 11.
-
Removable Chassis DX1000N
The DX1000N features a removable chassis via the front panel of the instrument. This allows easy access and maintenance of internal components of the DX1000N.