I set up time setting on the µR10000/µR20000 remote control function. I input a contact, but the time does not set. What should I check?

The time window in which you can input the contact and then set the time is limited.
While the time on the µR is between 2 m 00 s and 57 m 59 s, the time is not changed. Input the contact while the time on the µR is between 58 m 00 s and 1 m 59 s.

Related Products & Solutions

  • µR10000 & µR20000

    The μR series are the compact industrial recorders with the recording widths of 100 mm (μR10000) and 180 mm (μR20000). Measured data can be saved to an SD memory card while recording on chart paper.