The range of the strain input module depends on the measurement mode and gauge method. (see the table below)
Meausrement mode | 2K | 20K | 200K |
---|---|---|---|
1 gauge method | -2000 to +2000µε | -20000 to +20000µε | -200000 to +200000µε |
2 gauge method | -1000 to +1000µε | -10000 to +10000µε | -100000 to +100000µε |
4 gauge method | -500 to +500µε | -5000 to +5000µε | -50000 to +50000µε |
Even if you select the 2 gauge or 4 gauge method, 1 gauge method measurement is still possible, and rangeovers and other errors will not occur. However, the accuracies are only guaranteed within the ranges of each gauge method.