Information required in the CPMS plan includes identification of the specific flare being monitored, flare type, parameters to be monitored, and the expected parameter range including worst case and normal.
Installation Considerations
The sample gas pressure needs to be regulated to no higher than 70 psig.
The sample gas temperature needs to be no higher than 50°C (122°F).
A coalescing filter should be installed upstream of the detector to remove particulates and moisture.
The flow of the sample gas should be controlled at 0.6 LPM.
NOTE: Yokogawa can provide a sample flow panel for hydrogen cooled generators that will regulate the pressure and flow as well as filter the sample gas. (Part number M1233ZM)
Cleaning the Detector
The sample gas density detector is very resilient against the normal sources of contamination: dust and oil misting. If a process upset results in the detector being “slugged” with a large amount of oil or water, the detector can be cleaned on site and the measurement restored.
Flow 99% pure isopropyl alcohol through the detector to remove the contaminate. Do not use “rubbing alcohol” which can contain 30% water. This would result in an extended drying time.
Flow clean dry instrument air or nitrogen through the detector until the unit is dry (the reading becomes stable).
Calibrate the unit and return to service.
Spare Parts
The Yokogawa gas density analyzer system does not normally require spare parts. It has no consumable items. If the internal o-rings of the detector are damaged, we recommend that the unit be sent to an authorized Yokogawa service center for repair. If circumstances demand that the o-rings be replaced on-site, please contact the Yokogawa Technical Assistance Center at 1-800-524-7378 for instructions.
For the M1233ZM sample flow panel, we have spares available for the filter, pressure regulator, and flow meters.
Please see the illustrations below for parts identification.
GD40 Gas Density Detector
Item | Part Number |
---|---|
Inlet Screen | K9213FS |
O-Ring #1 | Y9010EV |
O-Ring #2, #3 | K9213MM (QTY OF 2) |
O-Ring #4 | Y9111XA |
O-Ring #5 | K9213MN |
O-Ring #6 | K9213MP |
M1233ZM
Sample Flow Panel for Hydrogen Cooled Generator Applications
Item | Part Number |
---|---|
Filter (Complete) | M1233ZR-O |
Filter Element | M1233ZR-02 |
Dual Scale Flowmeter | M1233ZQ |
Pressure Regulator | M1233ZP |
Industries
-
Oil & Gas
Yokogawa has a wealth of experience in every part of the oil and gas business, from offshore and onshore facilities to pipelines, terminals, and deepwater operations. We provide solutions that enhance safety, ensure accurate and reliable operation, and increase plant efficiency.
-
Oil & Gas Downstream
The oil & gas downstream industry has been facing an increasing number of challenges in recent years. These include the changing characteristics of the feedstock to be processed, aging of process facilities and equipment, rising cost of energy, lack of skilled plant operators who can run a refinery safely and efficiently, and the ever-changing requirements from both the market and the customer.
Over the years, Yokogawa has partnered with many downstream companies to provide industrial solutions focused on solving these challenges and problems. Yokogawa's VigilantPlant solutions have helped plant owners to achieve maximum profitability and sustainable safety within their plants.
-
Oil & Gas Midstream
Yokogawa's innovative technology platforms and industry leading execution is well regarded in the global marketplace as a solutions partner pioneering the integration of technologies for all aspects of the oil & gas ecosystem from reservoir to enterprise. Proven solutions that include predicative business modeling, plant optimization and highly reliable automation platforms are assisting midstream operators to run their businesses at the optimal efficiencies. Yokogawa is helping clients to develop their automation roadmaps ensuring years of highly effective asset utilization and sustainability.
Related Products & Solutions
-
GD402/GD40 Gas Density Analyzer
The GD402 gas density analyzer and GD40 gas density detector provide continuous measurement of gas density as well as other valuable parameters, including specific gravity and percent concentration.
-
Analyzers
Process analyzers are used to determine the chemical composition and physical properties of substances used in industrial processes.
-
Gas Analyzers
Enhance efficiency and product quality with real-time analysis