I'm trying to receive settings from the recorder in the SMARTDAC+ standard hardware settings, but the E001 communication error message appears. What should I do?

Check the following items.

  1. Whether the IP addresses of the recorder and PC are set.
  2. Whether the LAN cable is correctly connected. In particular, whether it is connected to the recorder's internal LAN board rather than the GX90EX expansion module.
  3. Whether you are using the latest version of the SMARTDAC+ standard hardware settings.

 

相關產品&解決方案