Technically it is possible that when connecting a sensor to a powered adapter that one of the pins on the VP connector connects sooner than others.
The first step is to separate the sensor and adapter and put them back together again.
If that does not work then power down the SA11, connect the sensor and then power it up again.
The SA11 cannot be damaged.
Related Products & Solutions
-
All-in-One pH/ORP Sensor Series FU20 and FU24
The FU20 and FU24, all-in-one pH and ORP, sensors show how Yokogawa applies the motto "Simple is best" to sensor technology.
-
Digital SMART SENCOM™ Adapter, SA11
Reusable SMART adapter, requiring only the analog sensor to be disposed of when it reaches the end of its lifetime. With the SENCOM 4.0 platform, Yokogawa delivers reduced costs and waste while contributing to its long-term business goals of a sustainable future for all.