High Speed

Faster, Brighter, and More Versatile

The CSU-X1 is the high speed model of our CSU-series, which are widely recognized as the most powerful tools for live cell imaging.

  • World's fastest scanning speed of up to 2,000 fps
  • Microlens-enhanced Nipkow disk scanning
  • Filter wheel with six filter positions (in high-end model)
  • Exchangeable dichroic mirror block and emission filters
  • 3 configurations: 1-camera model, 2-camera model, bright field model
  • CO2 emissions reduced by 40%

LCA

Principles of the Microlens-enhanced Nipkow Disk Scanning Technology

A Nipkow spinning disk containing about 20,000 pinholes and a second spinning disk containing the same number of microlens to focus excitation laser light into each corresponding pinhole are mechanically fixed with a motor, and very rapidly raster scan the field of view with about 1,000 laser beams when rotated. The pinhole and microlens pattern are arranged in our proprietary design to optimize raster scan. Multi-beam scanning with the CSU-X1 not only increases scanning speed, but also results in significantly lower photobleaching and phototoxicity, because multiple excitation needs only a low level of laser power at the specimen to fully excite fluorescence.

nipkowdisk

Details

Faster

  • The world's fastest scanning speed (up to 2,000fps in full-frame).
  • Yokogawa's proprietary filter wheel with six filter positions moves to the adjacent position at 33 ms, world's fastest speed.

Brighter

Doubled*1 excitation power efficiency with newly developed beam shaper lens, allows use of lower power lasers, and could reduce camera exposure time.

efficiency
















Triplicated*1 S/N by cutting the background noise by one third, expands capability of really low-light imaging. Significantly brighter*1 images by using most efficient dichroic mirrors and emission filters.

Optical path comfiguration


More Versatile -2nd Camera Option*1

You can either simultaneously image two different emission ranges with two cameras, or can selectively use one of the two cameras you installed which is most suitable for your current experimental requirement. For each camera port, you can select to install high-speed filter wheel (Option). In addition to standard C-mount adapter, adapters for 8X8 EMCCD camera and F-mount camera are available (Option).

More Versatile -Second Camera Port Option


More Versatile -Bright Field Path Option*2

It allows you to use one camera for both confocal imaging with the CSU-X1 and bright-field (non-confocal) imaging through the bypass light path.

More Versatile -Bright Field Path Option


More Versatile—Exchangeable mirror block and exchangeable filters

Easily exchangeable dichroic mirror block and emission filters.

More Versatile—Exchangeable mirror block and exchangeable filters

*1 It is necessary to use your choice of commercially available dichroic mirrors for simultaneous multicolor imaging.

*2 The Bright Field Option is not applicable to some microscopy set-up due to steric interference. Please inquire the applicability.

High-end model (6-position filter wheel)

High-end model (6-position filter wheel)

High-end model (12-position filter wheel)

High-end model (12-position filter wheel)

Basic model

Basic model

  Main unit *1 Main unit +
Brightfield*2
Main unit +
Secondcamera*2
Confocal scanning method Microlens-enhanced Nipkow disk scanning
Scanning speed Choice:1,500~5,000rpm (Standard)
1,500~10,000rpm (High-speed)*3
External synchronization Scan-speed synchronization through pulse signals
Input : TTL level 300Hz ~ 2KHz
Corresponding to Nipkow disk spinnig speed
1,500~10,000rpm*3
Excitation wavelangth 405~647nm
Second port - Brightfield Second camera
Dichroic mirror Option*4
Dichroic mirror switching Automatic 3CH
(Dichroic mirror block can be exchangeable)
Laser beam input AFC-connector (Polished 8 degree)
Optical fiber Yokogawa's standard Single-mode polarization-preserving fiber*5
Filter wheel
(Emission side)
Filter wheel with six filter positions
Emission filter Option*4
Operation panel Switch open / close of laser shutter
External control RS-232C interface via Control unit
Microscope mount C -mount adapter
Operating environment 15~40℃ / 20~75% RH
Power consumption
(main unit)
24VDC 1A max.
Power consumption
(AC adapter)
Input:100 up to 240VAC±10%, 50 or 60Hz±3Hz, 75Wmax.
Output:24VDC 2.5Amax.
External dimension *6 175(W)×328.5(H)×301.5(L)mm 175(W)×328.5(H)×301.5(L)mm 175(W)×328.5(H)×301.5(L)mm
Weight*7 8.9kg 11.7kg 13.0kg

*1 Supplied with a control unit ( for filter wheel ) and a filter wheel.
*2 Supplied with two control units ( one each for filter wheel and for bright field ) and a filter wheel.
*3 Option.
*4 Filters are not included ( Excitation filter , Emission filter ,Dichroic mirror ) . Please inquire as you need.
*5 Supplied with each CSU-X1 main unit.
*6 Excluding protruding parts. Including filter wheel.

  Main unit *1 Main unit +
Brightfield*2
Main unit +
Secondcamera*2
Confocal scanning method Microlens-enhanced Nipkow disk scanning
Scanning speed Choice:1,500~5,000rpm (Standard)
1,500~10,000rpm (High-speed)*3
External synchronization Scan-speed synchronization through pulse signals
Input : TTL level 300Hz ~ 2KHz
Corresponding to Nipkow disk spinnig speed
1,500~10,000rpm*3
Excitation wavelangth 405~647nm
Second port - Brightfield Second camera
Dichroic mirror Option*4
Dichroic mirror switching Automatic 3CH
(Dichroic mirror block can be exchangeable)
Laser beam input AFC-connector (Polished 8 degree)
Optical fiber Yokogawa's standard Single-mode polarization-preserving fiber*5
Filter wheel
(Emission side)
Filter wheel with 12 filter positions
Emission filter Option*4
Operation panel Switch open / close of laser shutter
External control RS-232C interface via Control unit
Microscope mount C -mount adapter
Operating environment 15~40℃ / 20~75% RH
Power consumption
(main unit)
24VDC 1A max.
Power consumption
(AC adapter)
Input:100 up to 240VAC±10%, 50 or 60Hz±3Hz, 75Wmax.
Output:24VDC 2.5Amax.
External dimension *6 258(W)×329.8(H)×213.4(L)mm 259(W)×374.3(H)×248(L)mm

309.8(W)×329.8(H)×392(L)mm

Weight*7 7.8kg 10.6kg 12.2kg

*1 Supplied with a control unit ( for filter wheel ) and a filter wheel.
*2 Supplied with two control units ( one each for filter wheel and for bright field ) and a filter wheel.
*3 Option.
*4 Filters are not included ( Excitation filter , Emission filter ,Dichroic mirror ) . Please inquire as you need.
*5 Supplied with each CSU-X1 main unit.
*6 Excluding protruding parts. Including filter wheel.

  Main unit *1 Main unit +
Brightfield*2
Main unit +
Secondcamera*2
Confocal scanning method Microlens-enhanced Nipkow disk scanning
Scanning speed Choice:1,500~5,000rpm (Standard)
1,500~5,000rpm (High-speed)*3*4
1,500~10,000rpm (High-speed)*3*4
External synchronization Option*4
Excitation wavelangth 405~647nm
Second port

-

Brightfield Second camera
Dichroic mirror Option*5
Dichroic mirror switching Manual 1CH
(Dichroic mirror block can be exchangeable)
Laser beam input AFC-connector (Polished 8 degree)
Optical fiber Yokogawa's standard Single-mode polarization-preserving fiber*6
Filter wheel
(Emission side)
-
Emission filter Option*5
Operation panel Switch open / close of laser shutter
External control -*4
Microscope mount C -mount adapter
Operating environment 15~40℃ / 20~75% RH
Power consumption
(main unit)
24VDC 1A max.
Power consumption
(AC adapter)
Input:100~240 VAC ±10% ,50 or 60Hz 、75W max.
Output : 24VDC 2.5A max.
External dimension *6 175(W)×328.5(H)×213.4(L) mm 259(W)×373(H)×213.4(L)mm

308.5(W)×328.5(H)×213.4(L)mm

Weight*7 7.5kg 10.0kg 10.0kg

*1 Supplied with a control unit ( for filter wheel ) and a filter wheel.
*2 Supplied with two control units ( one each for filter wheel and for bright field ) and a filter wheel.
*3 Option.
*4 Requires control unit for rotation speed control and external synchronization.
*5 Filters are not included ( Excitation filter , Emission filter ,Dichroic mirror ) . Please inquire as you need.
*6 Supplied with each CSU-X1 main unit.
*7 Excluding protruding parts.

Control unit・Filter Wheel

Control unit
Type for 6-position filter wheel (F1) for 12-position filter wheel (F2) for bright field (B1)
Operating condition 15~40℃ 20~75%RH
Power consumption Input : 100 up to 240VAC ± 10%, 50 or 60Hz, 200VAmax
External dimension
(mm)
213(W)×132(H)×438(L) mm
Weight( 5.2 kg 5.2 kg 5.1 kg

 

6-position filter wheel for CSU-X1
Operatimg condition 15~40℃ 20~75%RH
Power consumption -
External dimension 112(W)×226(D)×100(H)
Weight 1.9 kg

 

12-position filter wheel for CSU-X1
Operatimg condition 15~40℃ 20~75%RH
Power consumption -
External dimension 154(W)×98(D)×154(H)
Weight 2.4 kg

Comparison between target and referenced product

Comparison between target and referenced product

  Target Product: CSU-X1 Referenced product: CSU22+Filter Wheel
Energy (MJ) CO2Emissions (kg) NOxEmissions (g) SOxEmissions (g) Energy (MJ) CO2Emissions (kg) NOxEmissions (g) SOxEmissions (g)
Raw materials 2,335.5 132.5 337.1 130.1 2,610.0 148.0 374.9 147.1
Components 10,510.2 494.5 1,423.4 452.8 15,329.2 728.1 2,061.7 671.7
Machiningand assembly 1,542.9 68.7 214.1 56.8 2,259.8 100.7 313.6 83.1
Logistics 2,717.7 185.5 210.8 471.1 3,211.7 218.9 252.3 553.9
Consumption 36,713.5 1,635.9 5,094.8 1,351.5 66,087.0 2,944.8 9,171.0 1,126.4
Disposal -569.9 -31.9 -75.7 -33.6 -612.8 -34.5 -81.2 -37.4
Total 53,249.8 2,485.3 7,204.4 2,428.8 88,884.8 4,106.0 12,092.2 3,851.5
functional factor 40.1% 39.5% 40.4% 36.9%  

 Note: Value of the referenced product is calculated based on the functional factor. Functional factor:2.16

We post our information to the following SNSs.
Please follow us.

 

Yokogawa Life Science

•Twitter @Yokogawa_LS
•Facebook Yokogawa Life Science
•LinkedIn Yokogawa Life Science
•YouTube Life Science Yokogawa

 

Yokogawa's Official Social Media Account List

Social Media Account List



Nikon

Nikon Instruments Inc.  
(North and South America)

Nikon Instruments Europe B.V.  
(Europe)

Nikon Corporation  
(Asia except for Japan)
 


 

Resources

Overview:

To investigate interactive dynamics of the intracellular structures and organelles in the stomatal movement through live imaging technique, a CSU system was used to capture 3-dimensional images (XYZN) and time-laps images (XYT) of guard cells.

Overview:

Faster, Brighter, and More Versatile Confocal Scanner Unit

Industries:
Yokogawa Technical Report
2.2 MB
Overview:

List of Selected Publications : CSU-X1

Downloads

Videos

Overview:

YOKOGAWA proprietary Spinning Disk technology enables fast real-time confocal imaging for applications such as high-speed 3D and long-term live cell imaging. These quantifiable imaging analysis are essential tools for modern precision drug discovery.
 

Overview:

Over past 20 years, YOKOGAWA proprietary Spinning Disk Confocal technology has been widely used as an indispensable imaging tool among top researchers. The technology enables faster live-cell observation with clearer and less photo-bleaching imaging.

Overview:

YOKOGAWA will contribute to technology evolution particularly in measurement and analytical tools to help build a world where researchers will increasingly focus on insightful interpretation of data, and advancing Life Science to benefit humanity.

News

Looking for more information on our people, technology and solutions?

Contact Us

Top