When measuring voltage with a universal input module or DCV/TC input module, what is the limit value where an overrange occurs?

For voltage measurement, if 5% of range + range span is exceeded, an overrange occurs (±32767 is output). For temperature range, the +10°C measuring range is the upper limit.
Ex) For the ±50 V range (span of -50 to +50), the upper limit overrange is determined as: 50 V+100 V × 0.05 = 55 V.