RAGK and RAGL

The Rotameters RAGK and RAGL are specifically designed for the measurement of low flows of liquids and gases. Their main application field is in purging, analyzer and gas sampling systems, laboratories and machinery equipment.
Integrated needle valves are optionally available and are used for the fine adjustment of the flow. The special design of the RAGL makes it possible to replace its glass tube without removing the complete device from the installation.
These simple and affordable Rotameters have a very broad application range, large variety in cones and floats and decades of proven design.
The Rotameters RAGK and RAGL complete Yokogawa's Metal and Glass Rotameter Portfolio suitable for SIL applications and applications complying to the Machinery Directive.

Details

Features

  • Suitable for applications according to IEC 61508:2010 and ISO 13849-1
  • Space-saving design
  • Needle valves for fine adjustment of flow
  • Negligible pressure loss
  • Cost-effective solution

Specifications

Refer to the General Specification sheet located under the 'Downloads' tab for detailed specifications.

For Pipe Sizes 1/4 inch to 3/8 inch
Process Connections Threaded, Swagelok, cutting ring, nozzle
Measuring Range Water 20 °C: 2 ml/h - 600 l/h
Air 20 °C 1 bar abs.: 0.1 l/h - 6300 l/h
Wetted Parts Material Borosilicate glass, 1.4571 (AISI 316Ti) or Polypropylene
Various float materials available
Gaskets NBR, FKM, PTFE
Process Temperature -25 °C to + 130 °C
Process Pressure 16 bar
Ambient Temperature -25 °C to 80 °C
Measuring accuracy acc. VDI/VDE 3513 sheet 2 (qG=50 %) 1.6 % / 2.5 % / 4 % /
Hazardous Area Approvals ATEX, IECEx and NEPSI
Suitable for Applications acc. IEC 61508
and Machinery Directive ISO 13849-1
Yes
Additional Features Limit switches for low/high alarms
Fine control valves
Pressure controller

FlowConfigurator

FlowConfigurator

Welcome to the FlowConfigurator:
Selecting the flowmeter that best meets your specific needs and requirements.
The FlowConfigurator allows you to generate your data sheet and save your sizing projects for easy reuse.

FlowConfigurator

 

Resources

Overview:

With the increased interest in green energy and renewable fuel sources, Biodiesel has emerged as a viable alternative to standard petroleum based diesel and home heating oil. It has zero CO2 emissions and can be created from waste vegetable oil or rendered animal fat. The number of Biodiesel processing plants has grown to over 2500 in the US alone.

Overview:

One of the most widely used methods of monitoring / controlling liquid level in a tank is the use of Bubble Tubes with Pressure or Differential Pressure Transmitters. A small but uninterrupted flow of air or inert gas is forced down through a dip tube which extends to near the bottom of the tank. The back pressure of the introduced gas is a function of the liquid level or head in the tank. The pressure transmitter takes the back pressure and converts it to an analog signal that is sent to the control room. The plant operator will then be able to monitor the exact level in the tank.

Overview:

For nearly 100 years, ozone has been used all over the world for water sanitation. It is nature's strongest oxidizer and has a byproduct of pure oxygen. Chlorine, on the other hand, is a very effective, time-proven chemical, but is ineffective against hard-to-kill microorganisms such as cryptosporidium; it also has some unpleasant byproducts. Ozone is recognized among the strongest, fastest, commercially available disinfectants and oxidants for water treatment, Ozone is used in both the disinfection of drinking water and bacterial killer in waste water treatment. Ozone is used in a large percentage of Municipal water treatment facilities as well as Pharmaceutical companies where pure water is essential.

Downloads

Looking for more information on our people, technology and solutions?

Contact Us

Top