Providing reliable chlorine measurements in Water Treatment Plants since 1959
No.1 Process analyzer supplier to Japanese market*
"Reliable Chlorine monitoring and minimize Chlorine injection"
"Continuous automatic cleaning and achieved less maintenance time "
"Maintenance remotely for efficient process management"
Many of our customers face these challenges - Yokogawa Chlorine Analyzers resolve them.
The FC800D non-reagent free available chlorine analyzer adopts the polarographic method, and a unique rotating electrode provides a continuous online measurement of free available chlorine concentration.
The FC800D is part of the SENCOM4.0 platform family and can connect to the FLXA402T. Thanks to a smart unit equipped behind the electrode unit, FC800D gives you more insight into the ordinary operation.
The FC800D is suitable for water distribution measurements at pumping stations, water delivery points, schools, and public facilities.
*Based on a Yokogawa survey in 2019
Features
Reagent free measurement, Reduced maintenance due to rotating electrode
- Amperometric measurement of free chlorine (polarographic method)
- Measure Free chlorine continuously without reagent
- Unique rotation electrode method keeps sensor clean and reduces maintenance cycle
- Easy maintenance structure, modularized motor unit and diagnostics functions strongly support regular maintenance
- FC800D is equipped with Manual calibration and Auto Calibration, 2 points correction, Basic (reference) sensitivity calibration, and circuit open Zero calibration
- Remote range switching is available
Configuration
Unique rotating electrode method keeps sensor clean
Continuous automatic cleaning with beads realizes contamination-resistant and stable measurement
You can reduce maintenance time dramatically thanks to easy maintenance structures.
- Smart unit: Built-in Main board
- Electrode Unit: Built-in Motor, Rotating contact
- Beads Case: Houses cleaning beads
- Needle valve for flow rate adjustment
- Rotating electrode: Rotating electrode with beads keep the surface polished. This keeps the measurement electrode contaminant free providing long term stability.
- Reference electrode: Built-in temperature sensor for temperature compensation
Applications
Each process at treatment plant: from reservoir to treated water, distribution system
Water Purification Treatment
- Stabilization
- Flocculation
- Settling Basins
- Chlorination
- Clean water reservoir
- Distributing pump
- Distribution tank & Piping
Specifications
Measured Object | Free chlorine in tap water | |
---|---|---|
Measuring System | Amperometric method (Polarographic method) using rotating electrode | |
Measuring Range | Free chlorie 0-1 to 0-3 mg/L(-F1), Combined chlorine insentive version 0-1 to 0-3 mg/L(-C1), Free chlorie 0-5 mg/L (-F2) (*), Free chlorie 0-10 mg/L( -F3) (*) (*) As for "-F2" and "-F3", please refer to the General Specifications. |
|
Sample Condition (-F1, -C1) | Temperature | 0 to 50°C (Temperature compensation is effective within 0 to 40°C) |
pH | No pH fluctuation pH6.5 to 7.5 (pH6.0 to 8.0 :Chlorine value 1mg/L(-F1)) |
|
Conductivity | No conductivity fluctuation 100 to 300 μS/cm (100 to 500 μS/cm :Chlorine value 1mg/L(-F1)) |
|
Flow Rate | 0.1- 2.5 L/min | |
Pressure | 1-150 kPa | |
SS (Suspended Solid) | under 10 mg/L | |
Power Supply | 100 to 240 V DC, 50/60 Hz | |
Digital Communication | Modbus TCP/IP, Modbus RTU | |
Installation | Indoor (Outdoor use requires protection against rain.Direct sunlight must be avoided.) |
Performance (-F1 and -C1)
Linearity | ±5% of reading or ±0.05mg/L whichever is greater |
---|---|
Repeatability | ±1% of reading or ±0.01mg/L whichever is greater |
* As for "-F2" and "-F3", please refer to the General Specifications.
About OpreX
OpreX is the comprehensive brand for Yokogawa’s industrial automation (IA) and control business and stands for excellence in the related technology and solutions. It consists of categories and families under each category. This product belongs to the OpreX Analyzers family that is aligned under the OpreX Measurement category.
Resources
SENCOM
Downloads
Instruction Manuals
- FLXA402T Liquid Analyzer for Turbidity and Chlorine Start-up and Safety Precautions (1.7 MB)
- FC800D Non-reagent Type Free Available Chlorine Sensor Unit Start-up and Safety Precautions (1005 KB)
- FC800D Non-reagent Type Free Available Chlorine Sensor Unit (6.5 MB)
- FLXA402T Setting Table (53 KB)
- FLXA402T Liquid Analyzer for Turbidity and Chlorine Operation of SC (1.2 MB)
- FLXA402T Liquid Analyzer for Turbidity and Chlorine Operation of pH (1.3 MB)
- FLXA402T Liquid Analyzer for Turbidity and Chlorine Operation of Converter (4.2 MB)
- FLXA402T Liquid Analyzer for Turbidity and Chlorine Installation and Wiring (Introduction and General Description) (3.9 MB)
- ST401G Sampling System (5.4 MB)
General Specifications
Technical Information
Drawings
MSDS/SDS
- K9041RA / Lubricating oil (221 KB)
- K9088PE / Polishing powder (Alumina) (166 KB)
Looking for more information on our people, technology and solutions?