There is no strict specification for the input current value of DCV/TC INPUT, but the current value can be determined from the following specification.
DCV/TC INPUT specification
- Max input voltage: V
- 2 VDC range, thermocouple, RTD: ±10 VDC
- 6 VDC range or higher: ±60 VDC
- Input resistance: R
- DC voltage 2 V range or less, or with thermocouples: 10 MΩ or higher
- DC voltage 6 V range or higher: Approx. 1 MΩ
Thus, the current value is (per Ohm's law, I=V/R)
- 2 V range or less, or with thermocouples: ±1 µA or less
- 6 V range or higher: Approx. ±60 µA
Be sure to check whether 60 µA or more is flowing, as this could damage the input module.
Related Products & Solutions
-
DA100-1
Model Name: DA100-1
-
DA100-2 Portable Hybrid Recorder
DA100-2 Portable Hybrid Recorder
-
DC100 Data Collectors
Model Name: DC100
-
DR130
The DR130 can record data in 10 colors from all measurement points at 2-second intervals. It has a variety of recording functions including analog trend recording over an effective recording width of 150 mm, recording of digital measured values, recording of various messages, zone recording, and partially compressed and expanded recording, thus enabling data to be read off speedily.
-