DAQ32 (DP120) or DAQWORX DAQ32 Plus (WX102) installed on Windows 7 or Windows Vista cannot change the IP address of DARWIN series Ethernet modules (suffix code /C7) or DARWIN series standalone models with the Ethernet communications option (model code DT300-41) shipped on or prior to June, 2007. Address this issue as follows.
- Method 1) Set the IP address from DAQ32 or DAQ32 Plus installed on Windows XP.
- Method 2) Set the IP address on the main unit menu (only possible on the DR/DC series).
- Method 3) Submit the unit for factory rework (charges apply) (contact your company's person in charge).