All Yokogawa analyzers use an open architecture with adjustable temperature compensation, adjustable isopotential point, and adjustable slope. They are compatible with all direct pH, ORP, and conductivity sensors.
There are two exceptions to this:
- Competitor digital SMART sensors will not work with Yokogawa analyzers. If the sensor has a built-in amplifier, it cannot be connected to a Yokogawa analyzer.
- Yokogawa inductive conductivity sensors must be used with Yokogawa analyzers and vice versa. A competitor inductive conductivity sensor cannot be used with a Yokogawa analyzer.
When connecting competitor sensors to a Yokogawa analyzer, the customer is responsible for determining all wiring connections given the wiring information available in the analyzer's instruction manual.
For additional assistance, please contact Technial Support at 1-800-524-7378 or support@us.yokogawa.com.