SC25F는 Yokogawa의 최초의 12mm 콤비네이션 디자인의 디지털 SMART pH 센서입니다. 일체형 온도 엘레먼트와 액체 접지 전극을 포함합니다.이 디자인은 대량의 전해액을 제공하여 대부분의 비교 가능한 pH 전극보다 센서의 수명이 깁니다.
Details
- Solid Titanium LE electrode for stable measurements
- Integral Pt1000 element for enhanced pH accuracy
- Versatile in-line, immersion or off-line installation
- CIP and Steam cleaning possible
- large internal KCL volume giving the sensor a longer life time
- low conductivity applications possible from 10µS
- SC25F-ALP25 for chemically harsh applications and high temperature
- SC25F-AGP25 for all General Purpose applications
- Easy setup by sensor specific characteristics stored in the sensor itself
- Simple maintenance by comprehensive design
- CE, ATEX, CSA and FM Approval. However *Can only be used General Purpose at this time, because FLXA21 SENCOM™ module and SPS24 Interface box are pending ICECx, ATEX, FM and CSA approval*
SPS24 PC Configuration and Calibration Software
WU11 Connection Cable
The WU11 interconnection cable is specified for reliable transfer of digital signals between the SENCOM™ sensors and Yokogawa FLXA™ analyzer. It is espically designed to be installed in a heavy industrial environment. The cable is water proof, IP67, allowing it to be submerged as a whole. The cable has a wide temperature operation range which is sufficient for most of the applications. The materials used for the cable are flame retardant to resist the spread of fire.
- Double shielding to protect connected devices for interference from high voltages and currents
- Available in four different lengths. Long cable runs up to 100 m (328 ft) are possible
- Easy connection to a SENCOM pH Sensor by one 5- pins molded connector, and to the analyzer by 5 wired pins
- Flame retardant recognized by UL (fileno. E124763) and CSA (fileno. LL105324)
- IP67
- Operating temperature from - 40ºC up to + 85°C (-40ºF up to +185°F)
자료실
Wet scrubbers are used in utilities, paper mills, and chemical plants to remove sulfur dioxide (SO2) and other pollutants from gas streams. Undesirable pollutants are removed by contacting the gases with an aqueous solution or slurry containing a sorbent. The most common sorbents are lime, Ca(OH)2, and limestone, CaCO3.
Optimizing four key factors will decrease pH sensor costs and optimize process control and overall plant efficiency.
Downloads
Instruction Manuals
General Specifications
Certificates
- EU_UK Declaration of Conformity SC25F 2022-11-14 (156 KB)
- EAC certificate for sensors (1.8 MB)
동영상
기술 및 솔루션, 요꼬가와에 대한 자세한 정보를 찾고 계십니까?
견적 및 기술문의