Measuring the Electric Conductivity by use of Clean-in-Place (CIP) System

Introduction

In the manufacturing process of Pharmaceutical, Chemical and Food & Beverage industries, the cleaning and sterilization of tanks and piping are done with various cleaning solutions, fresh or hot water and steam after manufacturing products. Clean- In-Place (CIP) is the system designed for automatic cleaning and disinfecting. The recovery of cleaning solutions contributes significantly to reduce chemical costs and wastewater processing costs.

Yokogawa FLXA202/FLXA21 inductive conductivity transmitter has earned a reputation in the recovery of cleaning solutions because it can take measurements with good boundary surface precision over a wide range.

CIP Process Flow

CIP Process Flow

Expected Benefits

  • Optimizes cleaning and sterilizing application for tanks and piping by use of CIP
  • Reduces operating costs

Process Overview

Inductive Conductivity SensorCIP systems depend on the process differentiations of Pharmaceutical, Chemical and Food & Beverage. In addition, the cleaning solutions and chemicals are different. The chemicals that are generally used are hypochlorous acid, caustic soda, and nitrous acid. In a cleaning solution recovery system, the process lines and piping are cleaned with an initial cleaning solution. After that, the cleaning solution is cleaned with fresh water. The cleaning solution is recovered at that point, but since it gradually becomes diluted with fresh water, recovery is discontinued and the solution is discharged through a drainage line when the concentration of the cleaning solution drops below the standard value. The cleaning with different solutions or fresh water, the recovery of the cleaning solution, and the discharge of waste solution are repeatedly carried out. In addition, sterilization with steam or cleaning with distilled water or de-ionized water is performed as a final step in some cases.

Solution Details

Measurement system
2-wire inductive conductivity system
Sensor ISC40G(J)
Flow-through type holder ISC40FF(J)
Transmitter FLXA202-D-B-D-☐☐-C5-NN-A-N-LA-J-NN
Dedicated distributor for transmitter PH201G-A☐*C

Utilities

FLXA202 inductive conductivity transmitter
Power supply voltage: 17 - 40 V DC (from distributor)
PH201G distributor
Power supply: 100 V: 20 to130 V DC/80 to 138 V AC, 47 to 63 Hz
220 V: 120 to 340 V DC/138 to 264 V AC, 47 to 63 Hz
Power consumption: 24 V DC: approx. 200 mA
100 V AC: approx. 7 VA
220 V AC: approx. 11 VA

Notes

Temperature range of transmitter is -20 to 140°C

Field Data

1. Process Conditions

  Concentration (%) Temperature (°C) Pressure MPa
Water - 4 to 20 0 to 1
NaOH solution 1 to 5 80 to 90 0 to 1
HNO3 solution 1 to 5 80 to 90 0 to 1
Steam - below 130 0 to 1

2. Features of ISC Inductive Conductivity Meter used in a CIP Process

  1. Improved precision in boundary surface measurement.
    (a) Equipped with a zero-suppression function according to output signal
    * Zero suppression
    Example:
    Output: 4 to 20 mA DC
    Range: 100 to 300 mS/cm
    (b) One sensor can handle a minimum span of 100µS/cm and a maximum span of 1999 mS/cm.
  2. Steam sterilization capability
    Measurements can be performed at process solution temperatures of -10 to 130°C and pressures no greater than 2 MPa {20 kgf/cm2}
Zero-suppression function
Zero-suppression function

업종

  • 식음료

    Yokogawa는 오늘날의 식음료 기업들이 기후 변화, 소비자 수요 및 글로벌 경쟁력 향상이라는 전례 없는 도전에 직면해 있다는 것을 알고 있습니다. 이러한 도전을 극복하기 위해서는 생산, 자산 관리, 식품 안전 및 품질이라는 핵심 분야에 초점을 맞춘 혁신적인 솔루션이 필요합니다.

    See More
  • 의약품

    제약회사는 제조 데이터를 기반으로 실시간 품질 관리를 강화하여 유연하고 민첩한 생산 체계를 구축해야 합니다. Yokogawa는 고객과 협력하여 안전하고 신뢰할 수 있는 의약품 제조 솔루션을 개발하고, 디지털 혁신을 통해 규제를 준수하고 품질을 보장하며 시장 출시를 앞당겨 환자에게 더 나은 의약품을 제공합니다.

    See More
  • 특수화학 및 정밀화학

    Yokogawa는 오랫동안 특수 및 정밀 화학 시장에서 고객을 지원해왔습니다. Batch 솔루션 설계의 복잡한 요구 사항을 이해하는 업계 전문가는 물론 업계 최고 수준의 안정성과 유연성을 제공하는 시장 선도 일괄 처리 솔루션을 사용하면 Yokogawa와의 파트너십에서 안전 및 규정 준수를 유지하면서 고객의 요구 사항을 충족시키는 제품을 생산할 수 있습니다.

    See More
  • 화학

    화학 플랜트는 연속 및 Batch 생산 공정에 의존하며, 각각은 제어 시스템에 대한 다양한 요구 사항을 제시합니다. 연속 공정은 실패하지 않고 생산 라인을 중단시키는 견고하고 안정적인 제어 시스템을 필요로하는 반면, Batch 공정의 중요성은 수식, 절차 및 공정을 조정하는 데 있어 큰 유연성을 허용하는 제어 시스템을 갖추는 데 있습니다. 두 종류의 시스템 모두 제품의 사용 가능한 품질 내역에서 관리되어야 하며 비일상적인 작업을 수행할 수 있어야 합니다. Yokogawa는 광범위한 제품 포트폴리오, 숙련된 시스템 엔지니어 및 글로벌 영업 및 서비스 네트워크를 통해 모든 공장 공정에 대한 솔루션을 제공합니다.

    See More

Related Products & Solutions

  • 2-Wire Transmitter/Analyzer FLXA202

    The FLEXA™ series analyzers are used for continuous on-line measurements in industrial installations. With an option for single or dual sensor measurement, they are the most flexible two-wire analyzer available.

    See More
  • Inductive (Torodial, Electrodeless) Conductivity Analyzer

    유도식 전도도 방법은 특히 높은 전도도 측정에 적합합니다. 분석계는 2개의 Troidal 변압기가 내장된 센서와 함께 작동합니다. AC 전류는 공정 샘플에 전압을 유도하여 샘플에 전류를 발생 시킵니다. 이 전류의 세기는 옴의 범칙에 따른 샘플의 전도도에 비례합니다. 

    See More
  • Conductivity Sensors

    전도도 센서와 전극은 공정 전도도, 저항률, WIFI, 탈염 수, RO 수, 농도 퍼센트, 보일러 블로우 다운 및 TDS를 측정하는 데 사용됩니다. Retractable, Flow-though, 침적 및 직접 삽입을 포함한 다양한 설치 옵션이 있습니다.

    적절한 전극 / 센서 선택은 최적의 측정 결과를 위해 중요합니다.

    See More
  • pH and ORP Analyzers

    pH와 ORP 미터, 분석계 및 전송기는 수질/제품 품질을 보장하고 유출물 배출량, 배치 중화, 펄프 스톡, 세정기, 냉각 타워, 화학, 폐수 처리 및 기타 여러 어플리케이션의 연속 공정 모니터링에 사용합니다.

    See More

Top