The FLXA402 can have maximum 4 outputs, what if a customer would like to have more than 4 outputs since we can connect five SMART sensors?

When using only analog sensors the maximum number of sensors connected are two. Then the maximum needed outputs are four (suffix code A4).
When using the SA11 Smart SENCOM adapter a maximum of 5 sensors can be connected. To be able to output all these measurements the ModBus RTU or ModBus TCP can be utilized.