What are the FC800D (Non-reagent Type Free Available Chlorine Sensor Unit) measurement ranges?

  -F1 -C1 -F2, -F3
Chlorin conc < 1 mg/L >1 mg/L
< = 3 mg/L
< = 3 mg/L < = 5 mg/L
< = 10 mg/L
pH pH: 6 to 8 pH: 6.5 to 7.5 pH: 6.5 to 7.5 pH: 6.5 to 7.5
Conductivity 100 to 500 μS/cm 100 to 300 μS/cm 100 to 300 μS/cm 200 to 300 μS/cm
Temperature Compensation 0 to 40 degree C 0 to 40 degree C 0 to 40 degree C 5 to 35 degree C

Related Products & Solutions


Top