RTD measurement with the BARD-700

Q:
With RTD input, the input source resistance is described in section 13.6, “General Specifications” of the user manual as: “10Ω or less per wire (the resistance of all three wires must be equal).” I'm using the BARD-700 safety barrier and the input resistance of the BARD is 130Ω, so can I perform measurement?

A:
Yes, you can perform RTD measurements while using the BARD-700 safety barrier. When using the BARD, make it so that the resistance value excluding the BARD is 10Ω or less. Make the resistance the same on each of the 3 wires.

Související produkty a řešení

  • Button Operated DX1000/DX2000

    The Daqstation DX1000/2000 series is a data acquisition system with display. Its HMI supports custom graphics for industry applications, audit trail and advanced security conforming to FDA 21CFR Part 11.

    Viz více
  • Removable Chassis DX1000N

    The DX1000N features a removable chassis via the front panel of the instrument. This allows easy access and maintenance of internal components of the DX1000N.

    Viz více

Nahoru