In the input range settings, you can add a specific value in the same manner as a bias setting by selecting a scale.
Ex) Set a thermocouple Type-K span of 0-100 DEGC. To display the temperature 0.5DEGC higher, set the following in the input range settings:
Mode = Scale; Type = TC; Range = K, Span lower limit = -0.5; Span upper limit = 99.5; Scale lower limit = 0.0; Scale upper limit = 100.0.