When I select the 1-5 V mode for the µR10000 range setting and enter a scale value I get Error 25. What should I do?

Range setting mode 1-5 V has a restriction in which you must set scale left < scale right.
Change to linear scaling, which does not have that restriction.

相关产品&解决方案


置顶
WeChat QR Code
横河电机(中国)有限公司