You can monitor the DX1000/DX1000N/DX2000 screen from Internet Explorer. Use the following steps.
- Set the IP address on the recorder and PC.
- In the recorder settings, under Basic Setting mode > Communication (Internet) > Server function, set Web to "Yes" (the initial setting is "Yes").
- In the recorder settings, under Basic Setting mode > Communication (Internet) > Web page, set Operator or Monitor to "ON" (initial value is OFF).
- (For Monitor page) In a PC web browser, in the address bar, enter http://[IP address of the recorder]/ and press the Enter key.
(For Operator page) In a PC web browser, in the address bar, enter http://[IP address of the recorder]/operator.htm and press the Enter key.
相关产品&解决方案
-
底盘可拆卸型DX1000N
DX1000N特点是仪器的前面板有一个可拆卸的底盘,这使得DX1000N的内部组件易于连结和维护。
-
按钮操作式DX1000/DX2000
Daqstation DX1000/2000系列是带有显示屏的数据采集系统。其HMI支持工业应用、审计跟踪和符合FDA 21CFR Part11高级安全性的自定义图形。