Evaluation of cell-culture condition in CV8000’s internal stage incubator

下载 (817 KB)

Introduction

Live-cell imaging is an extremely efficient technique that allows the collection of cellular timelapse information in a single image acquisition. The technique requires the preparation of devices that maintain an optimum environment (temperature, CO2 concentration, humidity control) for preserving cells in their normal condition. This technical note shows the evaluation results of the internal stage incubator of the CellVoyager CV8000 high-content analysis system. The evaluation consisted of the investigation of cell proliferation variation within well plates, and comparison against a conventional CO2 incubator.

Experimental procedure

A1 well B2 well

Fig 1(a). Timelapse movie: A1 well (72hrs)

Fig 1(b). Timelapse movie: B2 well (72hrs)

C3 well D6 well

Fig 1(c). Timelapse movie: C3 well (72hrs)

Fig 1(d). Timelapse movie: D6 well (72hrs)

  1. HeLa cells stably expressing Azami-Green (green fluorescent protein) were seeded (500 cells/well) in 250uL of medium in 96-well plates (Greiner #655896), and incubated for 24hrs. (Two plates were prepared.)
  2. One plate was incubated for 72 hours in a conventional CO2 incubator. Imaging was performed at 0hrs and 72hrs using the CV8000.
    The other plate underwent live cell imaging using the CV8000. Whole well imaging was performed at 3hr intervals for 72hrs with a 4x objective lens.
  3. Analysis was performed using the high-content analysis software CellPathfinder.
    The total area covered with cells in each well was calculated to evaluate the proliferation of cells. The variation of proliferation rate across wells was investigated. A comparison with the conventional CO2 incubator was also made.

Variation of cell proliferation rates within the well plate Fig 1(e). Line graphs of cell proliferation for each well

Vertical axis: total area (total area of cells in that well) Horizontal axis: time (0-72hrs)
Cell proliferation deteriorated in the plate’s corner wells; however, it continued at a good rate in the other wells.
Time-lapse movie under the similar condition: comparison of each wells:  Play

 

Fig 2. Variation of cell proliferation rates within the well plate (n=3)

Total area ratio for each well (total area at 72hrs / total area at 0hrs)(A) Total area ratio for each well (total area at 72hrs / total area at 0hrs)
96 well average: 7.7
96 well coefficient of variation (standard deviation/average): 8.5×10-2
36 well coefficient of variation (perimeter wells): 1.1×10-1
60 well coefficient of variation (except perimeter wells): 5.1×10-2

Change in total area ratio over time (error bars show standard deviation)(B) Change in total area ratio over time (error bars show standard deviation)
After 24hrs, there was no variation in the total area ratio across groups. Even after 72hrs, perimeter wells (except corner wells) showed similar results to the center wells.

Fig 3. Proliferation ratio comparison with a conventional CO2 incubator (n=3)

Proliferation ratio comparison with a conventional CO2 incubatorComparison of the cell proliferation ratio 72hrs after seeding, for each well, between a regular CO2 incubator and the CV8000
CV8000 total area ratio / CO2 incubator total area ratio×100
(The nearer the value to 100, the more similar cell proliferation was between the CV8000 and the CO2 incubator.)
96 well average: 90
36 perimeter well average: 81
60 well average (except perimeter wells): 96

Discussion

The results showed there was little variation in cell proliferation across wells (except perimeter wells) in the CV8000’s internal incubator, and its performance was comparable to that of a conventional CO2 incubator. It is generally understood that cell proliferation rates deteriorate on the perimeter of well plates due to culture medium evaporation. In this evaluation, however, a large variation in proliferation rates between the perimeter wells (except corner wells) and center wells was not observed. These results indicate that the CV8000’s internal incubator possesses sufficient performance for live-cell imaging.

 


Yokogawa Life Science Twitter Yokogawa Life Science Facebook Yokogawa Life Science Linkedin Yokogawa Life Sciences YouTube

行业

  • 制药

    制药公司必须更好地使用和监控制造数据,实时改善质量管理,从而提高灵活性和敏捷性。制药负责人与横河电机合作,一起研发生产解决方案,提供安全可靠的药品。我们共同借助于数字化转型和制造进步,满足监管要求,确保质量,加快上市时间,从而稳定、可靠地为患者供应药品。

    更多

相关产品&解决方案

  • CellPathfinder

    CellPathfinder is designed for our HCA systems, CQ1 and the CellVoyager series. From beginners to experts, the analysis software lets you quantify subtle physiological changes and even label-free samples with various graph options.

    更多
  • CV8000高内涵筛选系统

    CellVoyager CV8000是先进的高内涵筛选系统。改进的内置培养箱可以让客户分析长期的活细胞反应。凭借其可扩展性、4个摄像头、5个激光器和可选的内置移液器,该系统允许日益复杂的分析开发和高内涵筛选。

    更多
  • 生命科学

    解锁生命本质

    横河电机灵敏、高精度的解决方案可实现快速准确的测量,支持突破性的研究,加快药物发现,并优化大规模生物生产。

    更多
  • 高内涵分析

    我们的高内涵分析系统利用强大的分析软件助力从基础科学到药物发现筛选的一系列研究应用。

    更多

置顶
WeChat QR Code
横河电机(中国)有限公司