Introduction
Wet scrubbers are used in utilities, paper mills, and chemical plants to remove sulfur dioxide (SO2) and other pollutants from gas streams. Undesirable pollutants are removed by contacting the gases with an aqueous solution or slurry containing a sorbent. The most common sorbents are lime, Ca(OH)2, and limestone, CaCO3. In limestone-gypsum flue gas desulfurization systems, the consumption of a desulfurization agent (lime) is controlled using online pH analyzers.
Limestone-gypsum flue gas desulfurization consists of two processes: absorption and oxidation.
Absorption Process
After fly ash removal, the flue gas is bubbled through the scrubber, and the slurry is added from above. The lime or limestone reacts with the SO2 in the flue gas to create insoluble calcium sulfite (CaSO3) as shown in the equations below.
Reaction formulas:
Limestone: CaCO3(s) + SO2(g) → CaSO3(s) + CO2(g)
Lime: Ca(OH)2(s) + SO2(g) → CaSO3(s) + H2O(l)
A pH sensor in a recirculating tank is used to control feed of the solid lime or limestone. Neither lime nor limestone dissolves well in water and therefore, both are pumped in slurry form to the scrubber tower. In this process, it is important to promote the reaction of calcium (Ca) and to achieve a target desulfurization rate. The absorption rate is affected by the pH, concentration, temperature of liquid, and the liquid/gas ratio.
When the pH of the absorbing solution is below 5, calcium hydrogen sulfite (Ca(HSO3)2) is generated. Ca(HSO3)2 is converted into sulfur dioxide (SO2) in a reversible reaction, thereby reducing the desulfurization rate. When the pH of the absorbing solution is 7 or higher, carbon dioxide (CO2) in the flue gas reacts with calcium (Ca) to produce calcium carbonate (CaCO3). This increases lime consumption. Since the pH has an influence on the subsequent oxidation process, the pH of the absorbing solution should be kept between 5 and 6.
Oxidation Process
Reaction formulas:
CaSO3(s) + 1⁄2 O2(g) + 2H2O(l) → CaSO4・2(H2O)(s)
The calcium sulfite may be further reacted with oxygen by oxidation to produce gypsum. This is done normally under a pressure of approximately 500 kPa and a temperature of 50 to 80 °C. The pH of the absorbing solution in the absorber is approximately 6, and normally maintained at around 4.
Application Problems
The calcium compounds produced in scrubbers tend to accumulate in the recirculation loops and can cause a buildup of scale. Scale on the spray nozzles affects the atomization of the water droplets and reduces the scrubbing efficiency. The lime can cause scaling on the pH electrode which affect the pH measurement. To ensure accurate measurements, frequent cleaning of the electrodes with acid is required, which adds to both maintenance workload and cost. So having an accurate reliable pH control is necessary not only to control the lime and limestone feed, but also to forestall the start of scaling, as it is much easier to prevent scaling than to remove it.
Solution
One solution in the past was to install and auto cleaning installation that would save on maintenance time and expense, to help extended the life of the probe. However since the cation reference sensor, SC24V, was released in 2010 end users have changed out either existing Yokogawa product or competitors product and replaced it with the revolutionary differential pH/ORP sensor.
Since the cation sensor, SC24V, has NO junction, there is NO path from the process to the internal element; so NO poisoning can occur. Since there is NO junction, there is NO plugging or coating problems to worry about and there is NO electrolyte depletion problem, because there is NO electrolyte. The sensor virtually eliminates problems caused by aging and pollution of the liquid junction. Customers will have less maintenance time and longer sensor life.
Product Recommendations
Transmitter | FLXA202/FLXA21 2-wire pH/ORP measurement system |
---|---|
FLXA402 4-wire pH/ORP measurement system | |
Sensor | Option 1: Cation Reference Sensor SC24V with desired Holder (Retractable, Flow-Thru, Insertion); Any holder that accepts a PG13.5 connection |
Option 2: PR10 Retractable with combination electrode; or PH87 Retractable assembly with the PH97/DP Sensor | |
Option 3: FU20/FU24 pH/ORP Combination electrode | |
Option 4: FF20 Flow-thru assembly with individual measure, reference and temperature electrodes with the automatic cleaning assembly, K1547PJ |
i.e. SC21C-AGC55 and SM60-T1; or SM21-ALP26 and SM60-T1
行业
相关产品&解决方案
-
2-Wire Transmitter/Analyzer FLXA202
The FLEXA™ series analyzers are used for continuous on-line measurements in industrial installations. With an option for single or dual sensor measurement, they are the most flexible two-wire analyzer available.
-
Flow/NPT Fittings FF20/FS20
Yokogawa has invested considerable design and development time in producing a full range of fittings with particular emphasis on designs that reduce installation and maintenance time and consequently save operation costs.
-
FLXA202/21两线制变送器/分析仪
FLEXA™系列分析仪用于工业设施中的连续在线测量。它是灵活的两线制分析仪,可选择单传感器或双传感器测量。
-
FLXA402多通道四线制分析仪
FLEXA™系列分析仪是模块化设计的分析仪,用于在工业设施中进行连续在线测量。可提供单传感器或多传感器测量。
-
FU20/FU24/PH20一体化pH/ORP (REDOX)传感器
PH20/FU20/FU24一体化pH/ORP传感器体现了横河电机如何将“简单即好”这句格言应用于传感器技术。
-
Hot-Tap Holder PR10
On-line measurements often present extra challenges, especially when routine maintenance is required. The PR10 is ideally suitable for applications where the sensors must be removed without interrupting or shutting down the process. Without any special tools the PR10 can be retracted safely from the process at pressures up to 5 bar (72 psi). Using the PR10 allows us to place any dissolved oxygen sensor that has a PG13.5 connenction into a retractable assembly.
-
PH71/PH72便携式pH/ORP计
PH71/PH72 pH/ORP仪表结构紧凑,易于使用,采用防滴设计-非常适合现场使用。具有“一键校准”和温度补偿功能。实验室级智能pH计–无论是价格还是尺寸都非常适合您的口袋。
-
工业pH/ORP电极
pH值测量回路的关键是电极系统。横河电机研发了多种电极,确保电极系统在极其恶劣的条件下也可以运行。
-
差分pH/ORP传感器
阳离子差分pH/ORP传感器用于传统传感器失效的复杂应用,包括盐水溶液以及电解过程和奶酪生产等多种应用的测量。
-
pH和ORP传感器
pH电极和传感器是进行pH值测量的传感部分,有伸缩式、流通式、浸入式和直插式等多种安装选项。选择合适的pH电极/传感器对优化测量结果至关重要。
-
pH和ORP分析仪
pH和ORP流量计、分析仪和变送器用于pH和ORP的连续过程监测,以确保水/产品质量,如监测污水排放、批量中和、浆料存储、洗涤塔,冷却塔、化学品、水/废水处理和许多其他应用。