FSA130 Magnetic Flowmeter / Vortex Flowmeter Verification Tool

Magnetic Flowmeter / Vortex Flowmeter Verification Tool provides the diagnosis, reporting and data management for the health check of ADMAG TI (Total Insight) Series AXG or AXW and CA Series with HART protocol and Vortex Flowmeter VY series with HART protocol. The Verification Tool runs on FieldMate (Versatile Device Management Wizard) with R3.04.20 or later, and Device Files R3.09.21 or later.

* Click here for details on the Mobile Field Device Management FieldMate

*FSA130 is the license key to activate the Verification Tool.

Functions for Magnetic Flowmeter

There are cases that a magnetic flowmeter does not perform normally by failure of magnetic, excitation or calculation circuit, or insulation deterioration of excitation coil or electrodes. The Verification Tool can diagnose the health check of device without demounting a flowmeter from the process line.

Applicable model

  • AXG/AXW/CA Series Integral type flowmeter with HART protocol
  • Combination of AXG remote sensor and AXG4A, AXG1A remote transmitter with HART protocol
  • Combination of AXW remote sensor and AXW4A, AXG1A remote transmitter with HART protocol

 

Functions for Vortex Flowmeter

There are cases that a vortex flowmeter does not perform normally by failure of sensor circuit, signal processing circuit, calculation circuit or accumulation of fluid in shedder bar, deterioration of sensor. FSA130 is the license key to activate the Verification Tool.

Applicable model

  • VY series Integral type flowmeter with HART protocol
  • Combination of VY series remote sensor and VY4A remote transmitter with HART protocol

 

System Configuration

Magnetic Flowmeter

Magnetic Flowmeter Diagram
Image Zoom

* CA500 Series Multi-Function Process Calibrator is a product of Yokogawa Test & Measurement Corporation.

Vortex Flowmeter

Vortex Flowmeter Diagram
Image Zoom

* CA500 Series Multi-Function Process Calibrator and MY600 Digital Insulation Tester are products of Yokogawa Test & Measurement Corporation.


About OpreX

OpreX is the comprehensive brand for Yokogawa’s industrial automation (IA) and control business and stands for excellence in the related technology and solutions. It consists of categories and families under each category. This product belongs to the OpreX Field Instruments family that is aligned under the OpreX Measurement category.

Details

Magnetic Flowmeter

Standard Verification

This mode checks the status of magnetic, excitation, and calculation circuit of the device. It also checks the status of alarm occurrence, alarm history, cable connection status for signal and excitation cable, display board and LCD, and physical appearance of the device. This mode is performed with the magnetic flowmeter mounted onto process line filled with fluid.

Enhanced Verification

This mode checks analog input/output, pulse output, transmitter accuracy, and insulation resistance of the coil and signal electrodes for further device diagnosis. This mode is performed with the magnetic flowmeter being demounted from process line. The necessary are AM012 calibrator for magnetic flowmeter to simulate the flow velocity signal, CA500 series Multi-fucntion Process Calibrator to measure current value, pulse count and also to output current signal, and MY600 insulation resistance tester to measure the coil and electrode resistance. The CA500 and MY600 can be replaced by other instrument which has equivalent function.

Vortex Flowmeter

Standard Verification

This mode checks the status of sensor element of the shedder bar. Moreover sensor,signal processing, and calculation circuit of the device. It also checks the status of alarm occurrence, alarm history,display board and LCD, and physical appearance of the device. This mode is performed with the vortex flowmeter mounted onto process line filled with fluid.

Enhanced Verification

This mode checks analog input/output, pulse output, status output. CA500 series Multi-function Process Calibrator or other equivalent function equipment can be used.

Remote Maintenance

This mode displays the vortex waveform after signal processing, analysis of frequency, trend of output signal sensitivity of shedder bar. It is also possible to supports condition based maintenance.

Report to be generated

Device information

Information on the device to be inspected, such as model code and serial number, is displayed.

Magnetic Flowmeter

Standard Verification

  • Circuit: Magnetic Circuit, Excitation Circuit, Calculation Circuit
  • Device status: Alarm Occurrence, Alarm History
  • Connection Status*1: Cable Connection Status (Connection status of signal and excitation cable)
  • Indicator Board Check (Display Board Check): Indicator Status (Display Status), LCD display (Visual check by customer on test mode)
  • Physical Appearance: Flow sensor, Transmitter (Visual check by customer)

*1: Except CA Series

Enhanced Verification

  • Analog Output (Check by using CA500 or equivalent)
  • Pulse Output (Check by using CA500 or equivalent)
  • Analog Input *1 (Check by using CA500 or equivalent)
  • Transmitter *1*2 (Check by using AM012)
  • Insulation Resistance Coil, Signal *1*2(Electrode) (Check by using MY600 or equivalent)

*1: Excluding CA Series
*2: For remote type only, the item can be checked

Magnetic Flowmeter Diagram
Image Zoom

Vortex Flowmeter

Standard Verification

  • Circuit: Sensor circuit, Signal processing circuit, Calculation circuit
  • Device Status: Alarm Occurrence, Alarm History
  • Display: LCD display (Visual check by customer on test mode)
  • Physical Appearance: Flow Sensor, Transmitter (Visual check by customer)

Enhanced Verification

  • Analog Output (Check by using CA500 or equivalent)
  • Pulse Output (Check by using CA500 or equivalent)
  • Stataus Output (Check by using CA500 or equivalent)
  • Analog Input (Check by using CA500 or equivalent)

Vortex Flowmeter Diagram
Image Zoom

Downloads

Möchten Sie weitere Informationen über unsere Mitarbeiter, Technologien und Lösungen?

Kontakt

Top