In-Situ Gas Analyzer TDLS8000

The new TDLS8000 houses all of the industry’s leading features in one robust device. The platform design is for in situ measurements, eliminating the need for sample extraction and conditioning. The non-contacting sensor allows for a variety of process types including corrosive, abrasive, and condensing.

The first generation platform has been proven in many others for the measurements of O2, CO, CH4, NH3, H2O and many more NIR absorbing gases. This second generation platform has improved reliability and ease of installation and maintenance while still meeting or exceeding designed application demands.

What Is the In-Situ Gas Analyzer TDLS8000?

An industry proven laser analyzer designed specifically to meet all of your requirements in one robust device that is easy to operate and maintain.

Features

  • SIL2 TruePeak combined with smart laser Technology
  • Intuitive touchscreen HMI
  • HART and Modbus TCP communications standard
  • 8-stage auto-gain adapts to difficult applications
  • Fully field repairable with 50 days of data and spectra storage
  • Compact design for one-man installation without sacrificing ruggedness
  • Area classification Zone2/Div2 or Zone1/Div1

What Are the Benefits of the In-Situ Gas Analyzer TDLS8000?

System Configuration

Standard System Configuration
HART communication available

Standard system configuration

System Configuration with HMI

System configuration with HMI

Multi Analyzer Configuration with Remote HMI
Up to 4 units connection available

Multi analyzer configuration

 

High Reliability

Auto gain graphsReference Cell
Internal reference cell in the laser module ensures peak locking during trace measurement.

Auto gain
Auto-gain enables wide signal ranges against dynamic variation of transmission.

Validation
Validation can be initiated manually, remotely, or automatically on a daily, weekly or monthly basis defined by the user.

SIL2 certified
IEC61508 SIL designed & approved, SIL2 capability for single analyzer use, SIL3 capability for dual analyzer use.

 

TruePeak

The TruePeak we can measure the area of the absorbance peak. This eliminates effects from changing background gases, allowing for simple pressure and temperature compensation.

TruePeak spectra wavelengthsTraditional TDL spectra wavelengths

 

Intuitive touchscreen HMI

Touch screen 7.5 inch color LCD on HMI

  • Makes it simple to operate
  • Gives all the information including trend graph and eliminate PC to maintenance
  • Can be remotely installed

Touch screen HMITrend graph setting

 

Mini Display
Optical transmission at both the ends for easy alignment

Mini display optical transmission
 

Details

Standard Specifications

Measurement object O2, CO, CO or CH4, CO2, CO + CO2, H2O, NH3, NH3 + H2O, H2S, HCl concentration in combustion exhaust gas and process gas
Measurement system Tunable diode laser spectroscopy
Measured components and range Measured component Min. range Max. range
O2 0-1% 0-25%
CO(ppm) 0-200 ppm 0-10,000 ppm
CO+CH4 CO 0-200 ppm 0-10,000 ppm
CH4 0-5 %
NH3 0-30 ppm 0-50,000 ppm
H2O (ppm) in non HC 0-30 ppm 0-30,000 ppm
H2O (ppm) in HC 0-30 ppm 0-30,000 ppm
CO (%) 0-20 % 0-50 %
CO (%) + CO2 (%) 0-30 % 0-100 %
 NH3 + H2O NH3 0-30 ppm 0-5,000 ppm
H2O 0-5 % 0-50 %
H2S 0-5 % 0-100 %
CO2 (%) High Range 0-1 % 0-5 %
CO2 (%) Extend.Range 0-30 % 0-50 %
H2O (%) 0-10 % 0-100 %
HCl 0-50 ppm 0-5,000 ppm
Output path length Optical distance between the laser unit and the sensor control unit Standard; 0.5 to 6 m, Max; 30 m
Output signal 2 points, 4 to 20 mA DC
Output types; Gas concentration, Transmission, Process gas temperature, Process gas pressure
Output range; 3.0 to 21.6 mA DC
Digital communication HART, Ethernet
Contact outputs 2 points, contact rating 24 V DC, 1 A
DO;
Function: Activate during Warning / Calibration / Validation / Warm up / Maintenance conditions
Fault;
Function: Activate during normal condition, not activate during fault condition or when the system power is off
Valve control output 2 points
Function; Activate calibration or validation solenoid valves for zero, span or validation gas
Output signal; 24 V DC, 500 mA Max. per terminal
Alarm Warning;
Gas concentration low/high, Transmission low, Process pressure low/high, Process temperature low/high, Validation required, Validation failure, Zero/Span calibration error, Non process alarm, External alarm
Fault;
Laser module temperature low/high, Laser temperature low/high, Detector signal high, Peak center out of range, Reference peak height low, Absorption too high, Transmission lost, Reference transmission low, Reference peak height high, Laser unit failure, Laser module error, File access error, E2PROM access error.
Contact inputs
(Digital input)
2 points
Function;
External alarm/Calibration start/Validation start/Stream switch
Contact specification;
Zero voltage contact input
Input signal; Open signal;
100 kΩ or more, Close signal; 200 Ω or less
Input signal
(Analog input)
2 points, 4 to 24 mA DC
Input types; Process gas temperature, Process gas pressure
Self-diagnostics Laser Unit temperature, Sensor Control Unit temperature, Laser temperature, Detector signal level, Memory read/write function, Peak locking condition
Calibration Calibration method; Zero/Span calibration
Calibration mode; Manual, Auto, Local initiate (HMI)
Validation Validation method; Up to 2 points
Validation mode; Manual, Auto, Local initiate (HMI)
Power supply 24 V DC ±10%
Warm-up time 5 min.
Protection degree IP66, NEMA Type 4X
Hazardous area classifications Division 1, Zone 1; Explosion-proof/ Flame-proof type; FM, cFM, ATEX, IECEx (Pending)
Division 2, Zone 2; Non-Incendive/Type n; FM, cFM, ATEX, IECEx
Process gas condition Process gas temperature; Maximum 1500℃
Process gas pressure; Max. 1 MPa, Min. 90 kPa
Dust in process gas; 20 g/m3 or less
Installation condition Ambient operating temperature; -20 to 55℃
Storage temperature; -30 to 70℃
Humidity; 0 to 95%RH at 40℃ (Non-condensing)
Mounting flange type; ASME B 16.5, DIN, JIS
Gas connections; 1/4 NPT or Rc1/4
Purge gas;
   Recommended purge gasses
   O2 analyzer; N2 (99.99% or greater, application dependent)
 H2O ppm analyzer; N2 (99.99% or greater with < 20 ppm H2O for feed to the optional dryer package)
CO, CO or CH4, CO2, CO+CO2, NH3, NH3+H2O, H2S, HCl;
   N2 (99.99% or greater, application dependent) or Instrument air  CO, NH3 analyzer: N2 (99.99% or greater, application dependent) or Instrument air
Purge gas flow rates;
   5-20 L/min for optic
   5-30 L/min for process window
Purge gas connections;
   1/4NPT (-G1, -C2, -D2, -C2, -D1, -C1),
   Rc1/4 (-G2, -S2, -E2, -J2, -E1, -J1)

Performance

Measurement gas Repeatability Linearity
O2 ± 1% reading or ± 0.01% O2, whichever is greater ± 1% F.S.
CO (ppm) ± 2% reading or ± 1 ppm CO, whichever is greater ± 1% F.S.
CO + CH4 CO ± 1% reading or ± 1 ppm CO, whichever is greater ± 2% F.S.
CH4 ± 4% reading or ± 0.02% CH4, whichever is greater ± 4% F.S.
NH3 ± 2% reading or ± 1 ppm NH3, whichever is greater ± 2% F.S.
H2O (ppm) in non
HC
± 2% reading or ± 0.1 ppm H2O, whichever is greater ± 1% F.S.
H2O (ppm) in HC ± 2% reading or ± 0.1 ppm H2O, whichever is greater ± 1% F.S.
CO (%) ± 1% reading or ± 0.01% CO, whichever is greater ± 1% F.S.
CO (%) + CO2 (%) CO ± 1% reading or ± 0.1% CO, whichever is greater ± 1% F.S.
CO2 ± 1% reading or ± 0.1% CO2, whichever is greater ± 1% F.S.
NH3 + H2O NH3 ± 2% reading or ± 1 ppm NH3, whichever is greater ± 2% F.S.
H2O ± 4% reading or ± 0.05% H2O, whichever is greater ± 2% F.S.
H2S ± 1% reading or ± 0.005% H2S, whichever is greater ± 1% F.S.
CO2 (%) High Range ± 1% reading or ± 0.005% CO2, whichever is greater ± 1% F.S.
CO2 (%) Extend. Range ± 1% reading or ± 0.02% CO2, whichever is greater ± 1% F.S.
H2O (%) ± 1% reading or ± 0.004% H2O, whichever is greater ± 1% F.S.
HCL ± 1% reading or ± 2.5 ppm H2O, whichever is greater ± 2% F.S.

YH8000 HMI Unit

The YH8000 is a HMI designed specifically for the Tunable Diode Laser Gas Analyzer, the TDLS8000.

  • Touchscreen 7.5 inch color LCD on HMI
  • Makes it simple to operate
  • Can be remotely installed
  • Up to 4 units connection available

Specification

Display Touchscreen 7.5 inch TFT color LCD panel, 640 x 480 (VGA)
Communication Ethernet; RJ-45 connector, Communication speed; 100 Mbps
Protection degree
of enclosure
IP65, NEMA Type 4X
Weight 4 kg
Mounting Analyzer mount (Front, left-side, right-side) with tilt function, Pipe mount or Panel mount
Cable Entries 1/2NPT or M20 x 2
Installation
conditions
Ambient operating temperature; -20 to 55℃
Storage temperature: -30 to 70℃
Humidity: 10 to 90%RH at 40℃ (Noncondensing)
Power Supply 24 V DC ±10%
Hazardous area
classifications
Division 2, Zone2: Non-Incendive/Type n; FM, cFM, ATEX, IECEx

The simple and robust TDLS8000 assures the reliable and less maintenance operation of the analysis.

Fired Heater Combustion Safety and Lifecycle Management

Yokogawa TDSL8000 and CO + CH4 measurements provide reliable information to achieve:

  • Combustion efficiency improvement
  • Safety improvement
  • Longer lifetime of the coils and coil hangers
  • Higher throughout of the process heating

Limited O2 Concentration

O2 Measurement for Safety and Process Monitoring & Control Yokogawa TDLS8000 02 analyzer achieves:

  • No sampling system operation
  • Fast response analysis
  • No interference analysis
  • Less maintenance operation

 

Resources

Overview:

Fired heaters are used for various processes in oil refining and petrochemical plants.

Overview:

Ensure minor issues in Vinyl Chloride Production can be identified before becoming serious problems with Yokogawa’s In-Situ Gas Analyzer TDLS8000 and Probe Type Analyzer TDLS8200.

Industries:
Overview:

In maintaining and managing industrial plants, monitoring waste water pH/ORP is both a legal obligation and an unavoidable necessity for protecting the environment. Monitoring without an attentive eye can lead to severe consequences.

Industries:
Overview:

Considering safety and environmental issues such as combustion efficiency and decreasing NOX and CO in exhaust gas, it has become important to control O2 concentration in garbage incineration processes.

Overview:

NH3 injection lowers NOx in stack flue gas, boosts dust collection efficiency, and prevents erosion. Precise measurement and control of NH3 are crucial to avoid excess use, which increases costs and causes odors. The TDLS8000 solves these issues with direct measurement for effective NH3 management.

Overview:

Electrolysis plants create hydrogen and chlorine from a brine solution. Chlorine gas generated from the anolyte of the electrolysis tank generally contains between 0.5 to 2.0 vol% H2O. The sample is then cooled and filtered to remove brine, subsequently coming out as wet chlorine gas. The wet gas is sent to a drying tower where it is treated with sulfuric acid to get moisture down to the ppm level. 

Industries:
Overview:

O2 measurement is used to safely recover flue gas containing a high concentration of CO from a converter furnace. With conventional paramagnetic oxygen analyzers, O2 concentrations are obtained through a sampling system to recovery flue gas. Therefore, there are problems with the maintenance and running costs of the sampling system, and time delays in measurement.

Industries:
Overview:

Spectrometric technology can assess many critical characteristics about products, but it has limits. It can be challenging to determine when the line has been crossed

Industries:
Overview:

With fired heaters, users hope to get greater efficiency and reduced emissions but often are disappointed. Given the number of fired heaters operating every day and their importance in the process industries, any improvements realized across the board will have huge impacts. More units can reach their potential with some simple changes in work practices and technology upgrades.

Industries:
Overview:

SABIC Turns to Yokogawa TDLS Analyzers to Enhance Safety Functions and Production

Downloads

Videos

Overview:

Yokogawa’s new TDLS8000 houses all of the industry’s leading features in one robust device. The platform design is for in situ measurements which negate the need for sample extraction and conditioning. The non-contacting sensor allows for a variety of process types including corrosive, abrasive and condensing. The first generation platform has been proven in many others for the measurements of O2, CO, CH4, NH3, H2O and many more NIR absorbing gases. This second generation platform has improved reliability and ease of installation and maintenance while still meeting or exceeding designed application demands.