Yes, the SENCOM Module in these sensors allows a Modbus Master to
connect using Modbus RTU protocol and access the following data:
- pH
- Temperature compensated pH
- ORP
- pH compensated ORP
- rH
- Temperature
- Junction resistance value
- Sensor details (Model, Serial Number, production date)
- Sensor calibration data (zero, slope, temperature offset)
- Sensor status signals (e.g. Glass impedance detection)
As shown in the diagram above, our pH is calculated by subtracting value B from value A. To do this the B value is stored inside the sensor and can be displayed independently at any time. Since ORP is simply the value of B, we can easily display both pH (A-B) and ORP (B) without any issues.