I set the measurement range on the µR10000/µR20000 to thermocouple. The display indicates Over. What should I do?

Check whether the thermocouple is disconnected or broken. Also, with no input connected, try shorting plus and minus. If the input terminal temperature is displayed, the problem is not with the µR.