Industry: Electrical and Electronics
Product : Zirconia Humidity Analyzer


Humidity Measurement in Tantalum Capacitor Pyrolysis Furnaces
Steam Pyrolysis Process in Tantalum Capacitor Manufacturing
zoom

In tantalum capacitor manufacturing, there is a steam pyrolysis process where tantalum pellets are decomposed by steam. Humidity control in this process is important to maintain product quality and improve yield. Stable humidity measurement at high temperatures of 200 to 400 °C is required. The ZR402G/HS Direct In Situ Zirconia High Temperature Humidity Analyzer is easy to maintain as it requires no sampling system and utilizes a long life sensor with fast response. It is widely used in humidity instrumentation applications and has been well received in the marketplace.
- Maintains the quality of the tantalum capacitor manufacturing process
- Ensures stable, continuous humidity measurement
- Reduces operating costs
A tantalum pellet with an electrode plate applied is immersed in a manganese nitrate solution. This dipping process allows the manganese nitrate to penetrate the porous part of the pellet. After dipping, the pellet is forwarded to the pyrolysis process where the manganese nitrate is decomposed under high temperature and humidity conditions to form a manganese dioxide coating by the reaction Mn(NO3)2 -> MnO2 + 2NO2. These steps are repeated until the coating is sufficient. To improve the coating uniformity and yield, the humidity is controlled.
Field Data
  Measurement System
Measurement System
Process conditions
Measurement point Inside the steam pyrolysis furnace
Temperature 200 to 400 °C
Pressure10 to 20 Pa
Humidity 25 to 80 vol%H2O
Dust None




Measurement system
DetectorZR22G-☐☐☐-S-☐-C-R-☐-E-A/SV
ConverterZR402G-☐-E-E-A/HS/☐
Standard gas unitZO21S-☐-E*A




Utilities
Power supply rated voltage: 100 to 240 V AC
operating voltage range: 85 to 264 V AC
rated frequency: 50/60 Hz
operating frequency range: 45 to 66 Hz
Power consumption approx. 100 VA (300 VA max.)
Notes on Installation
As the sensor inside a pyrolysis furnace is hot, it can be easily damaged if cooler water drips onto it. The sensor should be installed with the probe head pointing downward.

Download printer-friendly version (332KB)