| Industry | : Wastewater | | Product | : Dissolved Oxygen Analyzer |
|
|
Dissolved Oxygen Measurement in Continuous Aeration Sewage Treatment
|
|
When used to measure the dissolved oxygen in an aeration
tank, sensors tend to quickly become contaminated and need
frequent cleaning. A cost-effective solution to this problem
has finally been found with the development of the Yokogawa
DO402 Dissolved Oxygen Converter system. With this system,
the sensor is mounted in a float ball holder that protects it from
contamination and enables continuous maintenance-free use for
extended periods. |
|
| - |
Continuously and stably measures dissolved oxygen
concentration |
| - |
Reduces operating costs |
| - |
Eliminates the need for manual cleaning |
|
|
|
Sensor Diaphragm
|
|
Sewage that flows into a treatment plant is sent to a primary sedimentation tank to remove the garbage and sludge. Thereafter, it is
routed to an aeration tank where activated sludge (sludge containing bacteria, protozoa, and other microorganisms) is added and air is
blown into the mixture to agitate it. The microorganisms in the activated sludge thrive on the oxygen and cause the organic substances
in the sludge to proliferate and the generated solids to be precipitated.
A shortage of dissolved oxygen (DO) will kill the microorganisms, so the DO concentration in the aeration tank must be properly
controlled. A DO analyzer is used to continuously measure the DO concentration.
The spherical float stays relatively clean
because contaminants are washed off
by the surface-level flow of water. The
sensor diaphragm is kept relatively free of
contamination as it is positioned at the same
height as the top portion of the float.
|
 |
| Measurement System |
|
Utilities |
| Detector: |
DO30G-NN-50☐☐-PN |
|
Power supply: |
100 VAC ±15 %, 50/60 Hz |
| Float holders |
|
|
115 VAC ±15 %, 50/60 Hz |
| |
Angled type: |
PB350G-PV-25-NN |
|
|
230 VAC ±15 %, 50/60 Hz |
| |
Vertical type: |
PB360G-PV-☐5-NN |
|
Power consumption: max. 10 VA |
| Converter: |
DO402G-1-☐-E/☐ |
|
|
|
Terminal box:
(as needed) |
WTB10-DO3-NN-☐☐/☐ |
|
|
|
| |
With the WTB10 terminal box, the sensors
and the transmitter can be up to 50 meters apart. |
| Precautions |
| - |
The measuring point must be at a location where air bubbles
are not a frequent occurrence. |
| - |
To prevent the sensor diaphragm from being damaged by
solids in the sample liquid, the arm of an angled float holder
should be pointed in the downstream direction. |
|
|
|
DO Readings Data
|
|
No effects of contamination after four months of continuous operation
We conducted a test with two DO30G sensors that were equipped with the PB350G Float Ball Holder. During the test, no maintenance
was performed on one of the DO30G sensors, while the other DO30G sensor was cleaned every two weeks and calibrated monthly.
Over the entire duration of the test, the DO30G sensor that did not receive cleaning showed minimal contamination. Furthermore, its
measurements did not deviate appreciably from those of the sensor that was regularly maintained.
|