EJX115A Low Flow Transmitter |
| L Capsule |
M Capsule |
H Capsule |
| Water equivalent |
0.016 to 7.2 l/min |
0.022 to 23 l/min |
0.07 to 33 l/min |
| Air equivalent flow |
0.44 to 198 Nl/min |
0.63 to 635 Nl/min |
2 to 910 Nl/min |
| Differential pressure span |
1 to 10 kPa (2 to 40 inH2O) |
2 to 100 kPa (8 to 400 inH2O) |
20 to 210 kPa (80 to 840 inH2O) |
| Accuracy |
IFO |
±5% |
±5% |
±5% |
| DP |
±0.04% |
±0.04% |
±0.04% |
| Degrees of protection |
IP67, NEMA 4X, and JIS C0920 immersion proof |
| Certificates |
FM, CENELEC ATEX, CSA, TIIS, IECEx |
| Output |
4 to 20 mA DC or FOUNDATION™ fieldbus or PROFIBUS PA 2-wire system with digital communication |
| Supply voltage |
BRAIN and HART:10.5 to 42 V DC (10.5 to 30 V DC for Intrinsically safe type) Fieldbus:9 to 32 V DC (9 to 24 V DC for Entity model and 9 to 17.5 V DC for FISCO model) |
| Ambient temperature |
-40 to 85 deg C (-40 to 185 deg F) (general use type) -30 to 80 deg C (-22 to 176 deg F) (with integral indicator) |
| Process temperature |
-40 to 120 deg C (-40 to 248 deg F) (general use type) |
| Maximum working pressure |
16 MPa (2300 psi) |
| Mounting |
2-inch pipe mounting |
| Wetted parts material |
Capsule |
316L SST (Diaphragm material is Hastelloy C-276) |
| Manifold |
316 SST |
| Cover flange |
ASTM CF-8M |
| Housing |
Cast aluminum alloy or ASTM CF-8M stainless steel (optional) |